SERIES 8000

Environmental chambers, Climatic cabinets
|
|
Technical Specifications environmental chambers, climatic cabinets
Back to presentation of environmental chambers
| Temperature control |
|
KB 8400 F |
KB 8400 FL |
| Temperature variation (time) |
+/- °C |
0,1 |
0,2 |
| Temperature deviation (spatial) |
+/- °C |
0,3 |
0,3 |
| Readability/ Setability |
°C |
0,1 |
0,1 |
| Temperature range |
°C |
-2 - 70 |
-2 - 70 |
| Sensor Thermocouple |
|
Type K |
Type K |
| Controller |
|
PID |
PID |
| Display |
|
LCD |
LCD |
|
Adjustable alarm limits (visual and acoustic) |
°C |
Yes |
Yes |
| Humidity control |
|
KB 8400 F |
KB 8400 FL |
| Humidity variation (time) |
+/-%RH |
2 |
2 |
| Humidity deviation (spatial) |
+/-%RH |
2 |
2 |
| Readability/ Setability |
%RH |
0,1 |
0,1 |
| Readability |
%RH |
1 |
1 |
| Range setting |
%RH |
1 - 99 |
1 - 99 |
| Working range |
%RH |
15 - 96**** |
15 - 96**** |
| Sensor Capacitive |
|
Yes |
Yes |
| Controller |
|
PID |
PID |
| Display |
|
LCD |
LCD |
| Acoustic alarm limits (visual and acoustic) |
%RH |
Yes |
Yes |
| Water quality |
|
Destilled/Ionized |
Destilled/Ionized |
| Water reservoir |
litres |
18 |
18 |
| Automatic De-icing system |
|
Fixed |
Fixed |
| Light control |
|
KB 8400 F |
KB 8400 FL |
| Readability/ Setability |
% |
|
3 - 100 |
| Light intensity, in the middle |
Lux |
|
12000 |
| Light intensity, in both sides |
Lux |
|
25000 |
| Safety termostats |
|
KB 8400 F |
KB 8400 FL |
| Temperature variation (time) |
+/- °C |
3 |
3 |
| Sensor (thermocouple) |
|
Type K |
Type K |
| Automatic setting |
|
Yes |
Yes |
| Adjustable limits |
|
Yes |
Yes |
| Timer |
|
KB 8400 F |
KB 8400 FL |
| Timer (1-999 minutes or hours) |
|
Yes |
Yes |
| Real Time Program |
|
Yes |
Yes |
| Real time Backup |
|
>1 month |
>1 month |
| Accessories |
|
KB 8400 F |
KB 8400 FL |
| Printer Report Program |
|
Optional* |
Optional* |
| Serial Data Port |
RS 232 |
Yes |
Yes |
| 2x24 Characters LCd Display |
|
Yes |
Yes |
| Access Port 30 mm |
|
Optional |
Optional |
| Inspection window in door with cover |
|
Optional |
Optional |
| Castors, lockable |
|
Yes |
Yes |
| Fittings |
|
KB 8400 F |
KB 8400 FL |
| Water filling port |
|
1/2" |
1/2" |
| Drain port |
|
1/2" |
1/2" |
| Max water pressure |
Bar |
6 |
6 |
| Shelves |
|
KB 8400 F |
KB 8400 FL |
| Standard (max) |
|
3 (22) |
3 (22) |
| Dimensions w,d |
mm |
610 x 580 |
610 x 580 |
| Max load per shelf |
kg |
30 |
30 |
| Permitted total load |
kg |
120 |
120 |
| Power consumption |
|
KB 8400 F |
KB 8400 FL |
| Nominal power |
W |
1200 |
1800 |
| Nominal voltage |
V |
230, 1~ |
230, 1~ |
| Frequency |
Hz |
50/60 |
50/60
|
| Dimensions |
|
KB 8400 F |
KB 8400 FL |
| Exterior W,D,H |
mm |
830,720,1840 |
960,720,1840 |
| Interior w,d,h |
mm |
630,592,1073 |
630,592,1073 |
| Volume working chamber |
litres |
400 |
400 |
| Weights |
|
KB 8400 F |
KB 8400 FL |
| Net weight |
kg |
210 |
240 |
| Shipping weight |
kg |
230 |
250 |
| Shipping dimensions, W,D,H |
cm |
91,80,201 |
104,800,201 |
| Shipping volume |
dm3 |
1463 |
1672 |
* This option is stored in the system and can freely be tested for 30 days. A unique code is needed in order to make it available for permanent use.
*** Ice build-up may occur on the main cooler at long runs below 0°C.
**** Humidifier disabled below 4°C and above 55°C
These technical data are specified for an empty cabinet and ambient temperature of 23°C.
|
|