cooled incubator

Series 8000
|
|
Technical Specifications cooled incubator
Back to presentation of cooled incubator
| Temperature control
|
| KB 8182
| KB 8400
| | KB 8400 L
| Temperature variation (time)
| +/-°C
| 0,1
| 0,1
| 0,2 |
| Temperature deviation (spatial)
| +/-°C
| 0,2
| 0,2
| 0,3 |
| Readability/ Setability
| °C
| 0,1
| 0,1
| 0,1 |
| Temperature range ***
| °C
| -9,9 - 70
| -9,9 - 70
| -2,0 - 70 |
| Sensor thermocouple
|
| Type K
| Type K
| Type K |
| Controller
|
| PID
| PID
| PID |
| Display
|
| LCD
| LCD
| LCD |
|
Adjustable alarm limits (visual and acoustic) |
|
Yes |
Yes |
Yes |
| Safety thermostats
|
| KB 8182
| KB 8400
| KB 8400 L |
| Temperature variation (time)
| +/-°C
| 3
| 3
| 3 |
| Sensor thermocouple
|
| Type K
| Type K
| Type K |
| Automatic setting
|
| Yes
| Yes
| Yes |
| Adjustable limits |
|
Yes |
Yes |
Yes |
| Light control
|
| KB 8182
| KB 8400
| | KB 8400 L
| Readabitity/Setability
| %
|
|
| 0 - 100 |
| Light intencity in the middle
| Lux
|
|
| 12000 |
| Light intencity in both sides |
Lux |
|
|
25000 |
| Accessories |
|
KB 8182 |
KB 8400 |
KB 8400 L |
| Automatic de-icing system |
|
Adjustable** |
Adjustable** |
Adjustable** |
| Timer (1-999 minutes or hours) |
|
Yes |
Yes |
Yes |
| Real Time Program |
|
Optional* |
Optional* |
Optional* |
| Printer Report Program |
|
Optional* |
Optional* |
Ootional* |
| Serieal Data Port |
RS232 |
Yes |
Yes |
Yes |
| Inspection window in door |
|
Optional |
Optional |
Optional |
|
|
|
|
|
| Shelves
|
| KB 8182
| KB 8400
| KB 8400 L |
| Standard/ max
|
| 3 (14)
| 3 (22)
| 3 (22) |
| Dimensions w,d
| mm
| 500 x 450
| 610 x 580
| 610 x 580 |
| Max load per shelf
| kg
| 20
| 30
| 30 |
| Permitted total load
| kg
| 120
| 120
| 120 |
| Accessories
|
| KB 8182
| KB 8400
| KB 8400 L |
| Printer Report Program
|
| Optional*
| Optional*
| Optional* |
| 2 x 24 characters LCD Display
|
| Yes
| Yes
| Yes |
| Access Port 30 mm |
|
Optional |
Optional |
Optional |
| Inspection window in door with cover |
|
Optional |
Optional |
Optional |
| Castors, lockable |
|
Yes |
Yes |
Yes |
| Power consumption
|
| KB 8182
| KB 8400
| KB 8400 L |
| Nominal power
| W
| 950
| 950
| 1200 |
| Nominal voltage
| V
| 230, 1~
| 230, 1~
| 230, 1~ |
| Frequency
| Hz
| 50/60
| 50/60
| 50/60 |
| Dimensions
|
| KB 8182
| KB 8400
| KB 8400 L |
| Exterior W,D,H
| mm
| 680, 580, 1430
| 830, 720, 1840
| 960, 720, 1840 |
| Interior w,d,h
| mm
| 520, 451, 777
| 630, 592, 1073
| 630, 592, 1073 |
| Volume working chamber
| litres
| 182
| 400
| 400 |
| Weights
|
| KB 8182
| KB 8400
| KB 8400 L |
| Net weight
| kg
| 95
| 180
| 210 |
| Shipping weight
| kg
| 120
| 210
| 240 |
| Shipping dimensions W,D,H |
mm |
760, 660, 1600 |
910, 800, 2010 |
1040, 800, 2010 |
| Shipping volume
| dm3
| 803
| 1463
| 1672 |
* This option is stored in the system and can freely be tested for 30 days. A unique code is needed in order to make it available for permanent use.
** Temperature limit, de-icing durations and rate can be adjusted . At low temperatures, this system will effect the temperature control.
*** Ice built up may occur on the cooler at long runs below 0°C. With 100% light, min temp is +5 °C
These technical data are specified for an empty cabinet and ambient temperature of 23°C
|
|