• Automatic safety thermostat                                                           
  • 2 LCD displays - explanatory text boxes
  • Touch buttons
  • Software
  • Temperature range up to 100C
  • PID temperature controller
  • Timer
  • Data logging
  • Real time program
  • Two temperatures
  • Keypad with pin code
  • Potential free alarm contact
  • Same regulator for TS and B models
  • 10 speed fan
  • USB Outlet
  • Sterilisation cycle

SERIES 9000: LABORATORY INCUBATORS

Models: Temperature controlB 9025B 9051B 9130B 9420
Temperature variation (time) +/- ºC 0,2 0,2 0,2 0,2
Temperature deviation (spatial) +/- % 1 1 1,5 1,5
Readability/setability ºC 0,1/0,1 0/1-0,1 0,1-0,1 0,1-0,1
Temperature range   * ºC tb-99,9 tb-99,9 tb-99,9 tb-99,9
Sensor thermo-couple type K Yes Yes Yes Yes
Controller PID PID PID PID
Display LCD LCD LCD LCD
TimerB 9025B 9051B 9130B 9420
Hours/minutes 99h 59m 99h 59m 99h 59m 99h 59m
Real time program Yes Yes Yes Yes
SafetyB 9025B 9051B 9130B 9420
Alarm flashing / acoustic  Yes Yes  Yes  Yes 
Alarm limit setable    ºC  Yes Yes  Yes  Yes 
Alarm safety setting    ºC Yes Yes Yes Yes
FeaturesB 9025B 9051B 9130B 9420
Fan speed steps 1/10 1/10 1/10 1/10
Door gasket silicone Yes Yes Yes Yes
 Glass innerdoor  Yes  Yes  Yes  Yes 
 Data port, Serial    USB  Yes  Yes  Yes  Yes 
 Pot free alarm output Yes  Yes  Yes  Yes 
ShelvesB 9025TB 9051B 9130B 9420
Standard/max    pcs 2/7 2/8 3/16 3/23
Dimensions WxD    mm 346x235 434x325 534x325 602x570
Max load per shelf    kg 20 20 20 30
Permitted total load    kg 50 50 70 100
HeatingB 9025B 9051B 9130B 9420
Heating time to 37C    mins 27 30 33 30
Heat transfer at 70C    W 65 75 90 190
PowerB 9025B 9051B 9130B 9420
Nominal power    W 480 480 930 1430
Nominal voltage    V 230,1~ 230,1~ 230,1~ 230,1~
Frequency    Hz 50/60 50/60 50/60 50/60
DimensionsB 9025B 9051B 9130B 9420
Exterior WxDxH    mm 490x480x500 580x575x550 680x675x750 753x845x1360
Interior WxDxH    mm  350x255x300 440x345x350 540x455x555 610x600x1180
Volume    litres 25 51 130 420
Weights/volumeB 9025B 9051B 9130B9420
Net weight    kg 20 35 52 120
Shipping weight    kg  25 40 60 140
Shipping volume    dm3 0,21 350 558 1280
 
 
tb = 2°C above ambient
*Can be set at two temperatures daily and weekly

 

          

          

At what temperature and time should I operate a hot air sterilizer?

The critical process parameters for dry air sterilization are temperature and time. Temperatures below 160°C will not give a satisfactory result. Holding time at 160°C is 2 hours, alternative 180° and 30 minutes holding time.

Temperatures above 180°C can harm instruments over time.

The new 9000 series incubators can also be used to sterilise. 

I have limited space. Can I stack the cabinets?

The three smallest models in series 9000 are table models and can be stacked on top of each other. 

I  would like to inspect the goods without opening the door. What can you offer?

The two largest models can be provided with glass in the door. An internal light for better viewing of the goods can also be provided.

Can you deliver the cabinets with castors instead of adjustable feet?

The cooling incubators KB 8000 and climatic cabinets/environmental chambers KB 8400 F are already provided with castors, all of them lockable. Other models can also be provided with castors as an extra option.